SF6 vs Methane: Choosing an Optical Gas Imaging Camera

SF6 and methane require different optical gas imaging considerations. Start with the target gas and the required output, then verify the camera configuration and its operating limits.

A similar-looking camera may have a different purpose

Optical gas imaging relies on spectral compatibility between the camera and the target gas. Housing shape and pixel count do not establish that compatibility. FLIR’s OGI application guidance explains why a camera suited to one gas group may not visualise another.

For purchasing, write the gas name or chemical formula into the enquiry. If the process contains a mixture, identify the relevant compounds and what the inspection is meant to establish. “Gas detection camera” alone is too broad a specification.

RG630F and RG630C: the distinction that matters

Selected manufacturer-listed differences
Model Example target gas Spectral range
Raythink RG630F Sulfur hexafluoride (SF6) 10.3–10.7 µm
Raythink RG630C Methane (CH4) 7–8.5 µm

The manufacturer’s RG630F and RG630C pages list additional gases. The table above highlights the selection difference; it is not a complete compatibility list. Confirm the gas, mixture and exact equipment revision before purchase.

For an SF6 equipment inspection and a methane process inspection, begin with separate requirements. Do not substitute one camera for the other because a product photograph looks the same.

Define what “detection” means in your project

Different outputs need different acceptance criteria
Required output Question for the supplier
Gas-plume visualisation Can the selected gas be visualised under the intended scene and viewing conditions?
Event or alarm What triggers the event, what output is available and what limitations apply?
Concentration What units, supported gases, method, calibration and conditions apply?
Leak rate Does the specific configuration provide this output, and what validation and operating limits support it?

Raythink currently lists concentration-related functions for both models. That does not make a plume video by itself a verified ppm reading or a mass-flow measurement. If a quantitative result is required, obtain the applicable technical documentation and agree how that function will be checked. FLIR’s explanation of OGI quantification also shows why quantitative methods have defined operating inputs and limits; its product-specific thresholds must not be applied to unrelated equipment.

Describe the conditions around the gas

Passive OGI performance can depend on the scene, including gas-to-background thermal contrast and weather. A demonstration recorded in one setting is not a universal detection guarantee. Specify the working distance, background, indoor or outdoor use and relevant environmental conditions. Ask for a representative example with those conditions stated.

For an evaluation, agree the target gas and the result that must be demonstrated before considering the camera accepted. Include the lens and any processing or software configuration in that record. Keep footage linked to its actual model rather than reusing a visually impressive clip from another camera.

Check site requirements separately

If equipment will enter a classified hazardous area, identify the required approval and ask for the certificate and marking that apply to the exact supplied configuration. A general statement about industrial use or a protection claim on a web page is not sufficient evidence for that installation.

An OGI camera should also be assessed within the site’s wider gas-management procedures. It should not be assumed to replace every point detector, personal gas monitor or required safety instrument. Define the role of each device with the responsible site team.

An enquiry that leads to a useful quotation

  • Gas or mixture, application and the output required.
  • Working distance, target location, background and environmental conditions.
  • Handheld survey or fixed monitoring requirement.
  • Required certificates, documents and software outputs.
  • Quantity, destination, target delivery date and evaluation requirements.

Request a gas-specific review

Share the target gas and required output. We will use that brief to review the relevant model information and quotation requirements.